Maintenance 119
ISE Detergent Solution is automatically run through the ISE module as part of the
shutdown wash procedure (WASH2). Manually perform additional ISE washes
(described in the following procedure) more frequently, but do not exceed 3 times in 24
hours in the following circumstances:
Dialysis samples are run routinely
Veterinary samples are run routinely
The system is in use more than 8 hours a day in a high volume laboratory, which is
having persistent imprecision or Chloride bias issues that have not been resolved by
any other means
NOTE: Do not perform the ISE electrode wash more than 3 times per day (once as part of
the shutdown wash and twice on a per-shift basis). Pour fresh ISE Detergent into a cup,
not a tube, before each wash, from the CTT.
1. At the Menu Panel, select Maint., then select ISE Operation.
2. At the ISE Operation window, in the Period.wash area, select
OFF, then select Set.
3. In the Wash Electrode area, do the following:
a. Place a cup of the ISE Detergent in the identified position on the screen.
b. In the Container field, select the type of container for the was solution. The
recommended type of container is
6 : 2mlCUP/Adp.
4. In the Wash Electrode area, select
Execute, then select Yes when prompted.
5. At the ISE Operation window, in the Period.wash area, select
ON, then select Set.
6. Close the window, then select
Yes when prompted.
Recording ISE slopes
Once a day, record the slopes from a successful ISE calibration on the Maintenance Log.
The slopes are provided on the ISE Monitor, RBL/Calibration History, and RealTime
Monitor windows following a successful calibration.
Weekly maintenance
Performing the weekly wash
Materials required:
• 5% solution of Reagent Probe Wash 3
(REF 03164495, PN B01-4183-01)
• ISE Wash (REF 01307361,
PN B01-4174-01)
• Deionized water
Time: 38 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.