144 ADVIA 2400 Operator’s Guide
Restoring system files
1. Insert the CD or the DVD containing the backup files into the CD drive.
NOTES
• When restoring backed-up data files (in the Data subfolder under the A002
folder), select the
Delete Data Files checkbox at the ADVIA Backup window.
This deletes any current data files on the PC hard drive before the backed up
system and data files are restored.
• If the current data files are needed, perform a backup before restoring previous
files. The restore feature restores all files (system and/or data files) that were
previously backed up. If this is the case, close the ADVIA Backup window and
perform Backing up system files (see above).
2. At the Startup window, select
Back-up.
3. At the ADVIA Backup window, select
Restore a Backup Copy, then browse to the
source folder that contains the backup files to restore and select
Execute.
4. At the Restore confirmation window, select
OK.
If the disk contains all the backed up files required for the restore procedure, the copy
function begins.
5. If the backed up files are on more than one disk, select
Continue.
6. At the ADVIA Backup window, select
Exit, select Restore, then select OK.
7. At the ADVIA Backup window, select
Cancel.
8. Reboot the PC.
Replacing the SPP, RPP1, and RPP2 probes
Materials required:
• Probes:
DPP (without crash detection),
REF 003975051, PN 073-0223-01
DPP (with crash detection),
REF 02030495, PN 073-0611-01
SPP-(REF 03975051, PN 073-0223-01)
RPP1,2-(REF 0551684, PN 073-0224-01)
• Phillips screwdriver
• Pliers
• Lint-free towels
Time: 10 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
NOTE: Use this procedure to replace SPP and RPP probes not equipped with crash
detection. For dilution probes (DPP) equipped with crash detection, refer to Replacing
DPP probes - with crash detection.