Operating the System 73
Creating workorders at the analyzer
1. At the Menu Panel, select Request, then select Order Entry.
2. Select
Routine or Interr.
3. In the Posi.no. boxes, enter the sample position number.
4. In the Samp.no. box, enter the sample identification number.
5. Verify that the System Dilution Mode, Container Type, Sample Type, Dil. factor,
Sex, and Blood collection date entries are correct.
6. As needed, provide entries for Comment and Age.
7. Order tests by any of the following methods:
In the Test table, select each test or ratio you want to run.
In the Test-tbl no. box, enter the number of each test you want, then press the
period (.) key.
In the Profiles area, select each profile you want to run.
8. To de-select all tests, select
Delete Tests.
9. To confirm, select
Yes.
10. To erase all entries, select
Batch Func.
11. To confirm, select
Yes.
12. Select
Enter.
The Number of workorder box increments. Automatic incrementing is assigned using
the Entry Setup window. If autoincrement is on, a new workorder displays with the
next sample number and position number incremented.
13. You can create another workorder or select
Exit to leave.
If necessary, select
New to clear the window for entry of the next workorder.
Creating multiple workorders
1. At the Menu Panel, select Request then select Order Entry.
2. Enter information for the first sample:
NOTE: Enter only those items you want replicated.
a. Select New located above the Enter.
b. Select Routine or Interr.
c. In the Posi.no. boxes, enter the starting sample position number (Tray and Cup
numbers).
d. In the Samp.no. box, enter the starting sample identification number.
e. Verify that the System Dilution Mode, Container type, Sample type, Dil. factor,
Sex, and Blood collection date entries are correct.
f. As needed, provide entries for Comment and Age.
g. Order tests by any of the following methods: