EasyManua.ls Logo

Siemens ADVIA 2400 - Reaction Tray Mixers; Reaction Tray Wash Mechanisms

Siemens ADVIA 2400
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 ADVIA 2400 Operator’s Guide
The RRV contains 340 reusable cuvettes (20 sets of 17 cuvettes).
The RRV cuvettes are immersed in the reaction tank in an oil-heat bath. This keeps the
cuvettes at a constant temperature of 37°C for sample analysis.
Reaction tray mixers
1)
Mixer 1 (MIXR1)
2) Mixer 1 wash port
3) Mixer 2 (MIXR2)
4) Mixer 2 wash port
Figure 1-14 Reaction tray mixers
The reaction tray (RRV) mixers (MIXR1 and MIXR2) stir the contents (sample and
reagent) of RRV cuvettes brought to their respective mixer positions.
Mixing is performed using a spinning, reciprocating, and vibrating paddle. Strong and
weak stirring options are available.
Both mixers are located behind the reaction tray. MIXR1 mixes sample with reagent 1
(R1). MIXR2 mixes sample with reagent 2 (R2).
Reaction tray wash mechanisms
1)
Reaction washer (WUD)

Table of Contents