118 ADVIA 2400 Operator’s Guide
6. At the WASH Set window, define the WASH2 container positions as follows:
a. Select
WASH2.
b. Select
2 for Cycles.
c. Enter
49 in the CTT cup position 1st time field and 50 in the CTT cup position
2nd time field.
d. Enter
49 in the RTT1 and RTT2 cup positions 1st time fields and 50 in the RTT1
and RTT2 cup positions 2nd time fields.
e. Select
Execute.
Performing additional ISE electrode washes
Materials required:
• ISE Detergent Solution
(REF 01307361, PN B01-4174-01)
Time: 5 minutes
Analyzer mode: Manual Operation
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
CAUTION
Use only Siemens-recommended
ancillary reagents to avoid
erroneous test results.
WARNINGS
Read and follow the cautions on
the box and label before
handling any reagents. Some
method and system reagents are
classified as hazardous material.
Avoid contact of method or
system reagents with skin and
eyes. These materials can cause
infection and burns. Wear
suitable protective clothing,
gloves, and eye/face protection.
In case of contact with eyes,
rinse immediately with plenty
of water and seek medical
advice. In case of contact with
skin, immediately wash with
soap and water.
Avoid inhaling chemical
vapors. If you inhale chemical
vapors, promptly leave the area
and seek fresh air.