72 ADVIA 2400 Operator’s Guide
Shutdown
WASH2 + System End + Power OFF = 39 minutes
Automatic startup
PRIME1 + WASH3 = 28 minutes
PRIME1 +WASH3 + Cell blank (wait for CB Temp) = 72 minutes
PRIME1 + WASH3 + Cell blank (no wait for CB TEMP) = 42 minutes
Daily Operation
Using workorders
Using host computer workorders
1. Create the workorders at your host computer.
Each patient sample must have a workorder that contains a sample number and at
least one test request.
NOTE: Stop here if you do not want to download workorders manually from the host
computer before operation.
Instead, use the Automatic item select feature in the Automatic transfer area on the
Online Settings window.
2. Download workorders to the chemistry system:
a. At the Menu Panel, select
Request, then select Order Entry.
b. Select
Host request.
c. In the Entry format area, select the means for identifying the first workorder in
step 2e.
d. In the Last no. entry format area, select the means for identifying the last
workorder in step 2f.
e. In the Start no. box, identify the first workorder you want downloaded.
f. In the Last no. box, identify the last workorder you want downloaded or enter the
number of workorders you want downloaded.
g. Select Execute.
After the workorders are downloaded, you can manage them in the same way as the
system workorders.
If a workorder already exists for the sample, the host workorder is used. See
managing host and system workorders for details.