132 ADVIA 2400 Operator’s Guide
3. Immerse the cuvette segments in 5% Probe Wash 3 solution.
a. Ensure no air bubbles are in the cuvettes.
b. Allow the cuvettes to soak for at least 10 hours.
4. Wash the cuvettes under running water, then rinse them in deionized water.
5. Drain the water from the cuvettes.
6. Install the cuvette segments on the dilution tray (DTT) and fasten the thumbscrews
by hand.
7. At the Operation Panel, select Initialize.
NOTE: Verify that the Operating mode field displays READY before performing any
further actions.
8. At the Operation Panel, select WASH.
9. Perform the daily shutdown wash routine, then verify the operation.
Cleaning and replenishing the cuvette conditioner bottle
Materials required:
• Deionized water
• Cuvette conditioner
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
NOTE: The cuvette conditioner bottle may be cleaned when it is refilled, but must be
cleaned at least once every 2 months.
1) ISE buffer bottle
2) RRV (Reaction) bath oil
3) Isotonic saline diluent bottle
4) Cuvette detergent bottle
5) Cell conditioner bottle
Figure 5-20. Cuvette conditioner bottle
1. Open the filter cap at the front of the cuvette conditioner bottle (5), and pull up the
tube with the filter.
2. Disconnect the cuvette conditioner bottle level sensor connector.
3. Turn the connector counter-clockwise and pull it out.