Fault message Description Measures
Memory error detec‐
ted
A hardware fault, faulty 󹪝rmware or no 󹪝rmware at all has been de‐
tected in the following analyzer modules, e.g., AM1. Possible causes:
No 󹪝rmware is present on the analyzer module.
The existing 󹪝rmware is invalid.
The program memory is faulty.
In this case, the analyzer module starts in bootloader mode and is
ready to perform a 󹪝rmware update.
If the error still exists despite the 󹪝rmware update, the analyzer mod‐
ule FBG must be replaced.
Please contact the service de‐
partment.
Cable fault detected The installed analyzer modules could not be recognized correctly.
Possible causes: Faulty or incorrectly connected CAN bus cable.
Check the CAN bus connec‐
tions.
Maintenance and servicing
11.10Eliminating faults and errors
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 155