EasyManua.ls Logo

Siemens SIPROCESS GA700 - Page 155

Siemens SIPROCESS GA700
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fault message Description Measures
Memory error detec‐
ted
A hardware fault, faulty 󹪝rmware or no 󹪝rmware at all has been de‐
tected in the following analyzer modules, e.g., AM1. Possible causes:
No 󹪝rmware is present on the analyzer module.
The existing 󹪝rmware is invalid.
The program memory is faulty.
In this case, the analyzer module starts in bootloader mode and is
ready to perform a 󹪝rmware update.
If the error still exists despite the 󹪝rmware update, the analyzer mod‐
ule FBG must be replaced.
Please contact the service de‐
partment.
Cable fault detected The installed analyzer modules could not be recognized correctly.
Possible causes: Faulty or incorrectly connected CAN bus cable.
Check the CAN bus connec‐
tions.
Maintenance and servicing
11.10Eliminating faults and errors
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB 155

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Related product manuals