Modbus
register
Data type /
Size in bytes
Parameter Description Default value
[units]
(units register)
Value range /
Setting options
Access
level
1124 Float / 4 Calibration val‐
ue 14
Calibration correction
factor that is associ‐
ated to Calibration
point 14.
1.0 0.5 - 2.0 Read /
write
(Expert)
1126 Float / 4 Calibration
point 15
Calibration point 15 for
undirectional flow or
calibration point 15 for
bidirectional positive
flow.If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1128 Float / 4 Calibration val‐
ue 15
Calibration correction
factor that is associ‐
ated to Calibration
point 15.
1.0 0.5 - 2.0 Read /
write
(Expert)
1130 Float / 4 Calibration
point 16
Calibration point 16 for
undirectional flow or
calibration point 16 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1132 Float / 4 Calibration val‐
ue 16
Calibration correction
factor that is associ‐
ated to Calibration
point 16.
1.0 0.5 - 2.0 Read /
write
(Expert)
1134 Float / 4 Calibration
point 17
Calibration point 17 for
undirectional flow or
calibration point 17 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1136 Float / 4 Calibration val‐
ue 17
Calibration correction
factor that is associ‐
ated to Calibration
point 17.
1.0 0.5 - 2.0 Read /
write
(Expert)
1138 Float / 4 Calibration
point 18
Calibration point 18 for
undirectional flow or
calibration point 18 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1140 Float / 4 Calibration val‐
ue 18
Calibration correction
factor that is associ‐
ated to Calibration
point 18.
1.0 0.5 - 2.0 Read /
write
(Expert)
1142 Float / 4 Calibration
point 19
Calibration point 19 for
undirectional flow or
calibration point 19 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
Modbus communication
A.9 Modbus holding registers tables
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD 181