Modbus
register
Data type /
Size in bytes
Parameter Description Default value
[units]
(units register)
Value range /
Setting options
Access
level
1104 Float / 4 Calibration val‐
ue 9
Calibration correction
factor that is associ‐
ated to Calibration
point 9.
1.0 0.5 - 2.0 Read /
write
(Expert)
1106 Float / 4 Calibration
point 10
Calibration point 10 for
undirectional flow or
calibration point 10 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1108 Float / 4 Calibration val‐
ue 10
Calibration correction
factor that is associ‐
ated to Calibration
point 10.
1.0 0.5 - 2.0 Read /
write
(Expert)
1110 Float / 4 Calibration
point 11
Calibration point 11 for
undirectional flow or
calibration point 11 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1112 Float / 4 Calibration val‐
ue 11
Calibration correction
factor that is associ‐
ated to Calibration
point 11.
1.0 0.5 - 2.0 Read /
write
(Expert)
1114 Float / 4 Calibration
point 12
Calibration point 12 for
undirectional flow or
calibration point 12 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1116 Float / 4 Calibration val‐
ue 12
Calibration correction
factor that is associ‐
ated to Calibration
point 12.
1.0 0.5 - 2.0 Read /
write
(Expert)
1118 Float / 4 Calibration
point 13
Calibration point 13 for
undirectional flow or
calibration point 13 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1120 Float / 4 Calibration val‐
ue 13
Calibration correction
factor that is associ‐
ated to Calibration
point 13.
1.0 0.5 - 2.0 Read /
write
(Expert)
1122 Float / 4 Calibration
point 14
Calibration point 14 for
undirectional flow or
calibration point 14 for
bidirectional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
Modbus communication
A.9 Modbus holding registers tables
SITRANS FST030 (Modbus)
180 Operating Instructions, 08/2018, A5E37367739-AD