Modbus
register
Data type /
Size in bytes
Parameter Description Default value
[units]
(units register)
Value range /
Setting options
Access
level
1084 Float / 4 Calibration val‐
ue 4
Calibration correction
factor that is associ‐
ated to Calibration
point 4.
1.0 0.5 - 2.0 Read /
write
(Expert)
1086 Float / 4 Calibration
point 5
Calibration point 5 for
undirectional flow or
calibration point 5 for bi‐
directional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1088 Float / 4 Calibration val‐
ue 5
Calibration correction
factor that is associ‐
ated to Calibration
point 5.
1.0 0.5 - 2.0 Read /
write
(Expert)
1090 Float / 4 Calibration
point 6
Calibration point 6 for
undirectional flow or
calibration point 6 for bi‐
directional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1092 Float / 4 Calibration val‐
ue 6
Calibration correction
factor that is associ‐
ated to Calibration
point 6.
1.0 0.5 - 2.0 Read /
write
(Expert)
1094 Float / 4 Calibration
point 7
Calibration point 7 for
undirectional flow or
calibration point 7 for bi‐
directional positive flow.
If 0.0 the calibration
point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1096 Float / 4 Calibration val‐
ue 7
Calibration correction
factor that is associ‐
ated to Calibration
point 7.
1.0 0.5 - 2.0 Read /
write
(Expert)
1098 Float / 4 Calibration
point 8
Calibration point 8 for
undirectional flow or
calibration point 8 for bi‐
directional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
1100 Float / 4 Calibration val‐
ue 8
Calibration correction
factor that is associ‐
ated to Calibration
point 8.
1.0 0.5 - 2.0 Read /
write
(Expert)
1102 Float / 4 Calibration
point 9
Calibration point 9 for
undirectional flow or
calibration point 9 for bi‐
directional positive
flow. If 0.0 the calibra‐
tion point is not used.
0.0
[m
3
/s]
(7500)
0.0 - 100000.0 Read /
write
(Expert)
Modbus communication
A.9 Modbus holding registers tables
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD 179