What to do if Siemens Measuring Instruments display shows no flow and measured sound velocity is 0 m/s with alarm codes 14, 15 and 34?
- SSarah LeonardSep 6, 2025
If your Siemens Measuring Instruments display shows no flow and the measured sound velocity is 0 m/s, and you're seeing alarm codes 14, 15, and 34, there could be several reasons. First, ensure the pipe is completely filled with liquid and consider relocating sensors to a normally full section, avoiding the top or bottom of horizontal pipes. Check the sensor cables for breaks or incorrect connections, verifying continuity by evaluating the DC resistance. Also, verify the pipe dimensions, sensor selection, mounting, and spacing, ensuring ultrasonic couplant is applied. Finally, the sensors or frontend hardware might be damaged; verify their operation using a pipe simulator 'Test block'.