Functions
8
8.1 Diagnostics
In this function group you can view all analyzer data. The menu sequence in the following Fig.
shows all submenus which can be accessed from the "Analyzer status" menu. The arrows lead
by one menu item to the next lower menu level which is called by this menu item.
This display applies to analyzers without an H
2
S sensor and without a paramagnetic O
2
sensor.
The di󹪜erences when using analyzers with one of these sensors are described in section
Diagnostics: Diagnostics values (Page 105). Dashed lines apply to analyzers with UV module.
81
8141
Firmware no.:
S79610-G210-A900
Softversion:VV.VV.VV
from DD.MM.YY 3.0
8131
Ser. no. : XX-XXX
Revision : X
Order no. :
7MBXXXX-XXXXX-XXXX-X
811
8114
8113
8112
8111
814
Basic unit
UV module
8142
Firmware no.:
A5E12345678
Softversion:VV.VV.VV
from DD.MM.YY 4.4
8111a

813
Basic unit
UV module
Revision : AA
Ser. no. : XXXXX
Order no. :
A5E12345678
8132
Status
Diagnostic values
Fact. settings hard
Fact. settings soft
IR drift readings
XXX AUTOCAL
Deviation: XXXX.XX %
Max.allowed:XXX.XX %
Select component
Level 1 required
Level 1 required
End of messages
press ENTER to clear
next message with ->
Logbook/Faults
Maintenance request
AUTOCAL deviation
O2 sensor status
O2 Sensor status
Reading : XX.XX mV
Maint.Req.: 6.00 mV
Fault : 5.00 mV
Flow too low during
AUTOCAL
press ENTER to clear
next message with ->
No maintenance
request registered
press ENTER to clear
next message with ->
Figure 8-1 Overview of diagnostic parameters
This overview contains all parameters of an analyzer that has both an IR detector and a UV
module. If only one of these detectors is present, this overview begins at menu screen 812b (IR)
or menu screen 812c (UV). Parameters marked with a footnote (*) are contained here in the
display for the IR detector. The di󹪜erences in the display for the UV module are described there
(section Diagnostics: Diagnostic values: UV (Page 107)).
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 101

Table of Contents