)XQFWLRQIRU
6,(0(16VHUYLFH
CAL/Pump/Sync inputs
Binary inputs
XXXXX vpm NNNNNZV
XXXXX vpm NNNNNOFF
XXXXX vpm NNNNNSV
XXXXX % O2 OFF
Test of binary inputs
Press ESC to cancel
Active input:
NONE
Test of display
Test of keys
Test of flow switch
Test in progress
Press ESC to cancel
Test of keys
Time left: XX sec
Pressed key:
No key pressed
Test of flow sensor
Flow O.K.
Press ESC to cancel
Test of C/P/S inputs
Active input:
none
Press ESC to cancel
Optional board
not found!
Press ESC to return
Analog 1: XX.XX mA
Analog 2: XX.XX mA
Analog 3: XX.XX mA
Analog 4: XX.XX mA
Test of relays
Relay number: X
Condition: XXXXXXXXX
Press ESC to cancel
843
Test:displ./keys/flow
Inputs/outputs
Chopper/IR source
RAM monitor
Test analog outputs
Test of relays
Test ext. sol. valve
Test of inputs
Chopper: ON
IR source: ON
843242a
843242
84321
843
8432
8431
84311
84312
84313
843241
84322 84323
84324
8433
Figure 8-7 Overview of conguration of analyzer tests
8.4.1 Conguration: Inputs/outputs/pump
Analog outputs
Assign relays
Binary/Sync inputs
Pump ON/OFF modes
841
You can use this menu to assign certain functions to the follow‐
ing elements:
• Relays
• Inputs and outputs
Furthermore, the following settings can be made using this
menu:
• Synchronization of several analyzers
• Pump response with AUTOCAL and in measuring mode
Functions
8.4 Conguration
ULTRAMAT 23
136 Equipment Manual, 04/2022, A5E37100388-012