The connection lines to the relay outputs, binary inputs, and analog outputs must be
shielded.
The analog outputs are 󹪞oating, but have a common negative pole.
As a measure to suppress sparking across the relay contacts (e.g. limit relays), RC elements
must be connected as shown in the following 󹪝gure. Note that the RC element results in a
drop-out delay for an inductive component (e.g. solenoid valve). The RC element should be
sized according to the following rule of thumb:
R = R
L
/2; C = 4L/R
2
L
,
where R = 100 Ω and C = 200 nF are su󹪟cient.
You must use a non-polarized capacitor for the RC element.
6ROHQRLGYDOYH
'HYLFHVLGH
&RQQHFWLRQVLGH
24 V
R
C
Figure 5-2 Measure to suppress sparks on a relay contact
When operated with direct current, a spark suppression diode can be installed instead of the RC
element.
Connect the signal lines to the Sub-D plugs at the rear of the device.
Refer to the ELAN interface description (Order No. C79000-B5200-C176 German, C79000-
B5276-C176 English) for details on the interface cable.
5.3.2 Power connection
NOTICE
Incorrect power supply
Check before connecting that the existing supply voltage corresponds to that speci󹪝ed on the
label of the device.
Install the power line separately from the signal lines.
A power supply cable or an appliance plug is enclosed with the device, and must only be
connected by quali󹪝ed personnel (see Quali󹪝ed Personnel (Page 11)). The cable is connected to
the appliance socket at the rear of the device. At the power supply end, the cable is inserted into
a mains socket.
Connecting
5.3 Electrical connection
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 73