3.4.6 Hydrogen sulde sensor
The cyclic Autocal must be switched on, see also Application note (Page 163).
General information
Measuring ranges
• Smallest measuring range
• Largest measuring range
0 ... 5 vpm
0 ... 50 vpm
Service life of the sensor Approx. 12 months after rst use in the ULTRAMAT 23
Shelf life Maximum of 12 months after date of manufacture
Operation mode Continuous measurement between 0 and 12.5 vpm
Discontinuous measurement between 12.5 and 50 vpm
AUTOCAL Cycle time: 1 hour (see Application note (Page 163))
Climatic conditions (restrictions)
Permissible ambient temperature
• During operation
• During transportation and storage
+5 ... +40 °C (41 ... 104 °F)
-10 ... +55 °C (14 ... 131 °F)
Permissible ambient pressure 750 ... 1200 hPa absolute
Inuencing variables
Associated gases The hydrogen sulde sensor cannot be used if the associated gas contains
the following components:
• Compounds containing chlorine
• Compounds containing uorine
• Heavy metals
• Aerosols
• Alkaline components (e.g. NH
3
>5 mg/m
3
)
Interference gases 1360 vpm SO
2
result in a cross-interference of <20 vpm H
2
S,
180 vpm NO result in a cross-interference of <150 vpm H
2
S,
no cross-interference of CH
4
, CO
2
and H
2
(1000 vpm)
Drift < 1% per month
Temperature < 3%/10 K based on full-scale value
Air pressure < 0.2 % of measured value per 1% pressure change
Measuring behavior at analog output
Output signal noise < 2% of smallest measuring range with a damping constant of 30 s