8.4.2.3 Con󹪝guration: Special functions: ELAN/PROFIBUS/external interference
8423
ELAN parameters
ELAN ext. interfer.
PROFIBUS parameters
Const. interference
You can use this dialog to con󹪝gure the analyzer for use in an
ELAN or PROFIBUS network.
The adjacent menu display appears when you select the function
from the higher-level menu.
Con󹪝guration: Special functions: ELAN/PROFIBUS/external interference: ELAN parameters
84231
Channel : 1
Name :ULTTWENTYTHREE
Protocol :OFF
Sync.zero cal. :OFF
You can use this dialog to set the parameters for an ELAN net‐
work. These are:
Channel
You must set the channel addresses for the analyzers present
in an ELAN network. Addresses from 1 to 12 can be set, where
each address may only be used once.
Name
You can set an analyzer name here. During communication
with ELAN, it can be used for the plain text identi󹪝cation of the
analyzers. An analyzer name may have up to 10 alphanumeric
characters.
Protocol (ON/OFF)
The automatic transmission of measured values can be switch‐
ed on/o󹪜. With 'ON', the analyzer sends a measured value
frame cyclically every 500 ms.
Note
To avoid considerably hindering communication within an
ELAN network, this function should only be set to 'ON' when
required (e.g. with correction of cross-interference).
Sync. zero cal. (ON/OFF)
This function is not yet available. Therefore only 'OFF' is the
currently valid value.
Note
For further details on operation of the analyzer in an ELAN network, please refer to the ELAN
interface description (C79000-B5274-C176 German/English).
Con󹪝guration: Special functions: ELAN/PROFIBUS/external interference: ELAN external
interference
84232
Interference CO2
Interferences : OFF
Interference 1
Interference 2
This function can be used for a correction calculation by measur‐
ing the in󹪞uence of an interfering gas by means of another ana‐
lyzer connected in the ELAN network.
Functions
8.4 Con󹪝guration
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 147