8142
Firmware no.:
A5E40715586
Softversion:VV.VV.VV
from DD.MM.YY X.X
If the device has a UV module, the information about this rm‐
ware can be called up in the additional menu screen (8142).
8.2 Calibration
In this function group you can use one or more calibration gases to calibrate the IR channels of
the ULTRAMAT 23 and to readjust the zero point and sensitivity (calibrate the full-scale value).
In addition, you can calibrate the oxygen sensor and the internal atmospheric pressure sensor
and also dene the AUTOCAL parameters. The gure shown below indicates the menu sequence
for an analyzer without UV module and H
2
S sensor. The menu displays and their functions are
described in the following sections.
The calibration functions can only be addressed if you have enabled code level 1.
82
Calibr. IR channels
Calibr. O2 sensor
Cal. pressure sensor
AUTOCAL/drift values
Total cal.:OFF NNNNN
with AUTOCAL: no
Span 1: XXXXX mg/m
Span 2: XXXXX mg/m
822
O2 cal after install
Start O2 zero cal.
Start O2 span cal.
826
Setpoint: XXXX mbar
Reading: XXXX mbar
827
Thermo AUTOCAL: OFF
Drift values
Cycle time AUTOCAL
Purge time
8212
O2 cal after install
Date :DD.MM.YYYY
8221
O2 zero cal. with N2
XX.XX % O2
Nitrogen flow please
Correct with ENTER
8222
Select component
Component NNNNN
Set span gas values
Start cal. MR 1
Start cal. MR 2
821
O2 span calibr.
Setpoint MR 1+2
Start cal. MR 1+2
8223
Level 1 required
3
3
8211
Set sp: XXXXX mg/m
Reading: XXXXX mg/m
Connect span gas
3
3
AUTOCAL
Cycle time: XX hours
time left: XX.XX h
Purge time
XXXX seconds
8272
8273
Select component
: NNNNN 1
8271
82711
Drift NNNNN vpm
Zero: X
MR1: X.X MR2: X.X
Reset drift values
AUTOCAL O2 sens: YES
AUTOCAL with N2: NO
Zero calibr. O2
Span calibr. O2
823
Span calibration
Setpoint MR 1+2
Start cal. MR 1+2
8232
Zero calibration
XX.XX % O2
Nitrogen flow please
Correct with ENTER
8231
Functions
8.2 Calibration
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 113