H
2
S sensor
If the analyzer has an H
2
S sensor, the starting menu for the calibration functions is changed. The
following overview shows the menu sequence for the calibration functions of the H
2
S sensor.
The calibration functions of the H
2
S sensor are described in section Calibration: H2S sensor
(Page 120).
82a
Calibr. IR channels
Calibr. sensor
Cal. pressure sensor
AUTOCAL/drift val.
82413
MR Cal.
Set span gas value
Start cal. MR1+2
8241
H2S Cal after inst.
Date :DD.MM.YYYY
Enter TC zero
Enter TC span
824
Level 1 required
Cal. after install
Start zero cal.
Start MR cal.
Input TC parameters
Select sensor
O2 sensor
H2S sensor
82411
H2S zero cal.
XXXXX vpm H2S
Zero gas flow
Calibrate with ENTER
82412
82414




Figure 8-3 Overview: Calibration with H
2
S sensor
UV components
If the analyzer has a UV module, the starting menu for the calibration functions is changed. The
following overview shows the menu sequence for the calibration functions of the UV module.
The calibration functions of the UV module are described in section Calibration: Ultraviolet
components (Page 124).
Functions
8.2 Calibration
ULTRAMAT 23
114 Equipment Manual, 04/2022, A5E37100388-012