Depending on your analyzer version, the starting menu may be dierent.
IR/UV diagn. values
O2 diagnostic values
P and T values
Other diagn. values
If the analyzer has an IR detector and a UV module, the adjacent
gure appears. After selection of the 'IR/UV diag. values' param‐
eter, the following prompt appears:
812a
IR Diagnostic values
UV Diagnostic values
This prompt asks whether the diagnostic values apply to IR-ac‐
tive or UV-active components.
• IR diagnostic values: After the prompt, the menu switches to
section Analyzer status: Diagnostics values: IR (Page 106).
• UV diagnostic values: After the prompt, the menu switches to
section Diagnostics: Diagnostic values: UV (Page 107).
IR diagnostic values
Sensor diagn. values
P and T values
Other diagn. values
G
If the analyzer has an H
2
S sensor, the adjacent gure appears.
The following versions are possible following selection of the
'Probe diagnostic values' parameter in the 2nd line:
• Only H
2
S sensor: After selection of this item, the menu
switches to the diagnostic values of the H
2
S sensor
(section Analyzer status: Diagnostics values: H2S sensor
(Page 109)).
• H
2
S and O
2
sensors: A query is made for the component, and
a switch then made to the called sensor.
8.1.2.1 Analyzer status: Diagnostics values: IR
IR diagnostic values
O2 diagnostic values
P and T values
Other diagn. values
E
This menu screen appears in the following cases:
• The analyzer has no UV module.
• In response to the prompt for whether the diagnostics ap‐
plies to IR-active or UV-active components, IR-active compo‐
nents were selected.
You must rst identify the component to which the diagnostic
values are to apply.
After the <ENTER> key is pressed, the following selection menu
appears:
Lockin-ADC + temp.
V-ADCt
Raw values
In this submenu you can call up the diagnostic values of the
selected infrared component. These are:
Lockin-ADC SO2
X= 408399 Y= 103444
R= 444912 V= 444872
B= 100116 T= 41.0°C
81211
• ADU are voltage and signal values of the analog-to-digital
converter prior to temperature compensation.
Functions
8.1 Diagnostics
ULTRAMAT 23
106 Equipment Manual, 04/2022, A5E37100388-012