8.4.2.2 Con󹪝guration: Special functions: AUTOCAL deviation (only for IR components)
8422
Preset AUTOCAL drift
Quantity: 4
Limit : 6 %
You can use this dialog to de󹪝ne the conditions under which a
drift from a reference value of a previous AUTOCAL triggers a
maintenance request. The adjustable parameters are:
Quantity:
The number of AUTOCAL procedures up to setting of a new
reference value (in this example: 4),
Limit:
The largest possible value in % of the set measuring range. In
the case of autoranging, range 1 is assumed with max. 99% of
the full-scale value. The deviation from the last AUTOCAL ref‐
erence value must not exceed this value (in this example: 6%,
seeCalibration: AUTOCAL/drift values (Page 125)), otherwise
a maintenance request is triggered.
1. 2. 3. 4.
t
$872&$/
6HWWLQJDUHIHUHQFH
YDOXH
$872&$/
6HWWLQJDQHZ
UHIHUHQFHYDOXH
Number of AUTOCAL procedures until a reference is set again (set
number: 4)
The previous deviation is still displayed with the 4th AUTOCAL; at
the same time the value of the 4th AUTOCAL is used as the new
reference value.
If a maintenance request 'AUTOCAL deviation too large' is ac‐
knowledged, the values are reset during the next AUTOCAL, and
counting commences at 1 again.
The following parameters are set at the factory:
Number of AUTOCAL procedures up to setting of a new reference value: 10
Maximum deviation until a maintenance request is triggered 6%
Maximum deviation until an error is triggered: 10%
Note
This function only applies to IR components.