8.2.4.4 Calibration: H2S sensor: Enter TC parameters
You use this function to enter the temperature compensation parameters for calibration of the
zero point and sensitivity. You can read these parameters from the sensor.
To do this, navigate as follows: Calibration -> Please enter code -> Calibrate probe -> Choose
component -> H
2
S -> Calibrate TC parameters. The following screen appears:
Enter TC zero
Enter TC span
82414
Temperature compensation (TC) of zero point
A: -3.0817e+1
B: +2.2517e+0
C: -1.1050e-1
D: +2.8011e-3
82414a
To do this, proceed as follows:
1. Position the cursor on the 󹪝rst line
2. Then press the <ENTER> key.
The adjacent display appears.
You can now view the factors of the temperature compensation parameters for the zero point,
and change them if necessary.
Temperature compensation of sensitivity
Enter TC zero
Enter TC span
82414
Proceed as follows for this (from menu display 'Enter TC param‐
eters'):
1. Position the cursor on the second line.
2. Then press the <ENTER> key.
A: +4.2117e+0
B: -2.8547e-1
C: +5.5451e-3
D: -2.0077e-5
82414b
The adjacent display appears.
You can now view the factors of the TC parameters for the sensitivity, and change them if
necessary.