8.4.2.1 Con󹪝guration: Special functions: Changing the codes/language
8421
Code 1 : 111
Code 2 : 222
Language: italiano
In the 󹪝rst two lines of this dialog, you can change the codes of
the two code levels 1 and 2 (see also section Code levels
(Page 95)).
The factory settings for the two code levels are:
Code level 1: 111
Code level 2: 222
You can also reduce the number of code levels by assigning the
same code to both levels.
Letters and special characters can also be used for the code, in
addition to numbers.
The modi󹪝ed codes are e󹪜ective immediately. You should there‐
fore make a note of the changed codes and keep this at a safe
place.
In the third line of this dialog you can change the language of the
input dialogs. The analyzer is designed for the following languag‐
es:
German
English
Spanish
French
Italian
Polish
A change is immediately e󹪜ective when you leave this dialog.