Note
A set limit only triggers a relay contact if a relay has previously been assigned to the
corresponding limit signal (see section Con󹪝guration: Inputs/outputs/pump: Assign relays
(Page 140)). The limits are not updated:
During the 󹪝rst warm-up phase
During an AUTOCAL
During the message: Function control and analog output at 'Hold measured value' (see
section Con󹪝guration: Inputs/outputs/pump: Analog outputs (Page 137))
Limit 1 SO2
Lim. 1: 2000 mg/m
Alarm conc. :-----
8321
3
If you have selected "Limit 1" or "Limit 2", a menu display appears
in which you can enter the lower or upper limit for each com‐
ponent. You can de󹪝ne the value in the second line, and the
condition under which a contact is be triggered in the third
(Alarm at conc.:):
High: with upward violation
Low: with downward violation
- - - - -: no message
The limit value has a hysteresis of ≤ 2% of the set full-scale value.
8.3.3 Parameters: Limits: H2S sensor protection
832
Limit 1 H2S
Limit 2 H2S
Protec. sensor H2S
If 'H
2
S' is selected as the component in the higher-level menu
display, the 'H
2
S sensor protection' function is selected in addi‐
tion to the limits. The 'Limit' function is described in section Pa‐
rameters: Limits (Page 130).
8323
Protec. sensor H2S :
ON
In the 3rd line you can select the 'H
2
S sensor protection' function.
The adjacent menu display appears when you select the func‐
tion in which you can switch the 'H
2
S sensor protection' function
on and o󹪜.
Hydrogen sul󹪝de (H
2
S) is a corrosive gas, especially in wet con‐
dition and in combination with other gases. This function pre‐
vents damage occurring to the H
2
S sensor from high concentra‐
tions of H
2
S. This function is enabled by default (ON). Execution
of this function is described in section Probe protection function
(Page 157).
Functions
8.3 Parameter
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 131