To do this, navigate as follows: Calibration -> Please enter code -> Calibrate probe -> Choose
component -> H
2
S -> Calibrate zero.
H2S Zero cal.
3.3 vpm H2S
Nitrogen flow please
Correct with ENTER
82412
The adjacent display appears.
In the case of analyzers with an internal sample gas pump, se‐
lecting this function switches from the sample gas inlet (inlet 1)
to the zero gas inlet (inlet 3).
Calibrate the zero point as follows:
Connect zero gas to the analyzer and observe the display.
Wait until the displayed value has stabilized.
Commence with correction of the zero point by pressing the
<ENTER> key.
Exit the calibration by pressing the <ESC> key.
The message "Tolerance not o.k." appears if an error occurs during the calibration.
8.2.4.3 Calibration: H2S sensor: Calibrating the measuring range
You can enter the setpoint and calibrate the sensitivity of the sensor using this function.
Navigate as follows to select this function: Calibration -> Please enter code -> Calibrate sensor
-> Select component -> H
2
S -> Calibrate MR.
MR Cal.
Set span gas value
Start Cal. MR 1+2
82413
The adjacent display appears.
Proceed as follows to enter the setpoint:
1. Position the cursor on the 2nd line of the display (Setpoint
MR 1+2)
2. Press the <ENTER> key
82413a
MB 1+2 10.00 vpm
The adjacent display appears. You can now enter the setpoint of
the calibration gas:
MR Cal.
Set span gas value
Start Cal. MR 1+2
82413
82413b
Set sp.: 50 vpm
Reading: 41 vpm
Connect span gas
Calibrate the sensitivity as follows:
1. Position the cursor on the 3rd line of the display
(Start cal. MR 1+2).
2. Press the <ENTER> key.
The adjacent display appears. Now carry out the following steps
for the sensitivity calibration:
1. Connect the sample gas inlet to the calibration gas.
2. Inject calibration gas to the sensor with a 󹪞ow rate of
1.2 ... 2.0 l/min.
3. Wait until the measured value has stabilized.
4. Then press the <ENTER> key.
5. Exit the calibration by pressing the <ESC> key.
The message "Tolerance not o.k." appears if an error occurs during the calibration.
Functions
8.2 Calibration
ULTRAMAT 23
122 Equipment Manual, 04/2022, A5E37100388-012