EasyManua.ls Logo

Siemens ULTRAMAT 23 Series - Page 111

Siemens ULTRAMAT 23 Series
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
81244
NO 3.11 4 /20mA
CO 4.25 4 /20mA
SO2 4.04 4 /20mA
O2 20.02 4 /20 mA
Analog output
For each gas component to be measured, the current value of
the output current in mA is shown (the unit is not shown for
space reasons) as well as the start-of-scale value (either 0, 2
or 4 mA) and the full-scale value (20 mA) of the output cur‐
rent range. For information on setting the start-of-scale val‐
ues, see section Con󹪝guration: Inputs/outputs/pump: Analog
outputs (Page 137).
81242
Reference : 4.99 V
Bridge : 3.379 V
Reference/bridge
Reference: The reference voltage for the electronics of
the analyzer.
Bridge: The supply voltage to the measuring bridge.
Version with UV module without IR analyzer section
D
Unlim. conc. values
Power/IR Source
This menu is used to call further diagnostic functions. You can
call the following values:
18 mg/m NO M1
2 mg/m CO M1
11 mg/m SO2 M1
20.77 % O2 M2
81243
3
3
3
Unlimited concentration values
Measured values that are not limited by the large measuring
range are displayed here. Negative measured values can also
be output here.
81241a
Power (VAC): 101 %
Raw voltage: 30.0 V
Power/IR source
Power: Data on the supply voltage in % of the nominal
value of the respective supply voltage (e.g.: 100% corre‐
sponds to 230 V or 120 V).
Raw voltage: This is the raw voltage after recti󹪝cation of
the transformer voltage
Note
All menu screens are available for IR/UV devices
Functions
8.1 Diagnostics
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 111

Table of Contents

Related product manuals