Siemens ULTRAMAT 23 Series - Page 142

Siemens ULTRAMAT 23 Series
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
84123
Fault :R-
Maintenace req.:R5
CAL/MEAS :R6
FunctionControl:R-
Status messages
You can use this menu to apply the signaling of various operat‐
ing states of the analyzer as an event for relay control
(R in the display means Relay).
The following signals are possible:
Fault:
Occurrence of a fault and output of a fault message
Maintenance request:
Occurrence of maintenance request
(assigned to relay 5 in Fig.)
CAL/MEAS:
Switching over from measuring mode to AUTOCAL (applied
to relay 6 in Fig.)
Service switch:
Occurrence of a function control.
External valves
zero gas : relay -
span gas : relay 7
Sync : relay 8
84124
In this menu you can trigger external solenoid valves via relay
contacts:
Zero gas:
The zero gas supply which is triggered with AUTOCAL
Span gas:
The calibration gas supply (assigned to relay 7 in Fig.)
Sync:
Synchronization of an AUTOCAL with other devices within a
system (assigned to relay 8 in Fig.; see section System setup
with several analyzers in parallel (Page 84)).
Functions
8.4 Con󹪝guration
ULTRAMAT 23
142 Equipment Manual, 04/2022, A5E37100388-012