Con󹪝guration: Special functions: Factory data/reset/units: Reset
84242
Warning! Analyzer is
set to warm-up mode!
Press ENTER to cont.
Press ESC to cancel
You can use this function to restart the analyzer.. When you select
this function, you will be warned (see
adjacent display) that the analyzer initially runs through a warm-
up phase following the restart and is thus unavailable for meas‐
urements for a certain time.
The <ENTER> key initiates the restart with the warm-up phase.
Triggering of the restart can be prevented here using the <ESC>
key.
Con󹪝guration: Special functions: Factory data/reset/units: Change units
84243a
Unit for SO2
set: mg/m
MR min: 400 - 10%
MR max: 2000 + 10%
In the second line of this menu display, you can change the fac‐
tory-set units of the measured components.
After changing the unit, the display of the 'MB min' and 'MB max'
parameters is adapted accordingly.
This dialog is speci󹪝c to the component.
84243a
Unit for SO2
set : vpm
MR min: 148 - 3%
MR max: 757 + 3%
Note
The full-scale values may assume unusual values as a result of the component-speci󹪝c
conversion factors. Subsequent adaptations can be carried out as described in
section Parameters: Measuring ranges: Setting measuring ranges (Page 129). In addition, you
should also check these parameters following this change:
Setpoints of the calibration gases (section Calibration: Infrared component: Set span gas
values (Page 116))
Settings for the limits (section Parameters: Limits (Page 130)).