Part No. Designation Remarks
1 *) IR source with temperature sensor
3 *) Chopper 2-channel
4 Chopper plate glued 2-channel with threaded rods
5.1 O-Ring for analyzer chamber
5.2 Analyzer chamber
5.3 Analyzer chamber Without o-Ring
6.1 Gas 󹪝lter
6.2 Gas 󹪝lter
7.1,7.2 *
**)
Receiver chamber
8 **) Optical 󹪝lter with 󹪝lter support
*) Following replacement of this part, special work is required which can only be carried out by a Siemens certi󹪝ed service center
or by quali󹪝ed, specially trained personnel, for example temperature compensation, basic electronic adjustment, 3-point
calibration with calibration gas etc.
**) Following replacement of this part, the water vapor cross-sensitivity must be checked.
Spare parts/accessories
13.6 IR analyzer units
ULTRAMAT 23
204 Equipment Manual, 04/2022, A5E37100388-012