Part No. Designation Remarks
4 Chopper plate glued 2-channel with threaded rods and windows
5.1 O-Ring for analyzer chamber required for analyzer chamber lengths 20 mm and more
5.2 Analyzer chamber
5.3 Analyzer chamber
7.1,7.2 *
**)
Receiver chamber
8 **) Optical 󹪝lter with 󹪝lter support
*) Following replacement of this part, special work is required which can only be carried out by a Siemens certi󹪝ed service center
or by quali󹪝ed, specially trained personnel, for example temperature compensation, basic electronic adjustment, 3-point
calibration with calibration gas etc.
**) Following replacement of this part, the water vapor cross-sensitivity must be checked.