7MB2338-, 7MB2358- (measured component and measuring range BA, BD, CB)
Part No. Designation Remarks
1 *) IR source with temperature sensor
3 *) Chopper 2-channel
4 Chopper plate glued 2-channel with threaded rods and windows
7MB2338-, 7MB2358- (measured component and measuring range BA)
Part No. Designation Remarks
5.2 Analyzer chamber Analyzer chamber 2 mm
7.1,7.2 *
**)
Receiver chamber
7MB2338-, 7MB2358- (measured component and measuring range BD)
Part No. Designation Remark
5.2 Analyzer chamber Analyzer chamber 1 mm
7.1,7.2 *
**)
Receiver chamber
7MB2338-, 7MB2358- (measured component and measuring range CB)
Part No. Designation Remarks
5.2 Analyzer chamber Analyzer chamber 6 mm
7.1,7.2 *
**)
Receiver chamber
*) Following replacement of this part, special work is required which can only be carried out by a Siemens certied service center
or by qualied, specially trained personnel, for example temperature compensation, basic electronic adjustment, 3-point
calibration with calibration gas etc.
**) Following replacement of this part, the water vapor cross-sensitivity must be checked.
Spare parts/accessories
13.6 IR analyzer units
ULTRAMAT 23
208 Equipment Manual, 04/2022, A5E37100388-012