EasyManua.ls Logo

Siemens ULTRAMAT 23 Series - Page 29

Siemens ULTRAMAT 23 Series
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Paramagnetic oxygen measurement
*DVLQOHW
*DVRXWOHW
Figure 3-6 Operating principle of the paramagnetic O
2
sensor
In contrast to other gases, oxygen is highly paramagnetic. This property is used as the basis for
this method of measurement.
Two permanent magnets generate an inhomogeneous magnetic 󹪝eld in the measuring cell. If
oxygen molecules 󹪞ow into the measuring cell (1), they are drawn into the magnetic 󹪝eld. This
results in the two diamagnetic hollow spheres (2) being displaced out of the magnetic 󹪝eld. This
rotary motion is recorded optically, and serves as the input variable for control of a
compensation 󹪞ow. This generates a torque opposite to the rotary motion around the two
hollow spheres by means of a wire loop (3). The compensation current is proportional to the
concentration of oxygen.
The calibration point is calibrated using the AUTOCAL function by connecting oxygen
(analogous to calibration of the electrochemical O
2
sensor). In order to comply with the
technical data, the zero point of the paramagnetic measuring cell must be calibrated with
nitrogen weekly in the case of all measuring ranges < 5% or every two months in the case of all
larger measuring ranges.
Description
3.3 Function
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 29

Table of Contents

Related product manuals