WARNING
Hazardous contact voltage
Risk of electric shock in case of incorrect connection.
For the electrical connection speci󹪝cations, refer to the information in Electrical connection
(Page 72).
At the mounting location of the device observe the applicable directives and laws for
installation of electrical power installations with rated voltages below 1000 V.
WARNING
Missing PE/ground connection
Risk of electric shock.
Depending on the device version, connect the power supply as follows:
Power plug: Ensure that the used socket has a PE/ground conductor connection. Check that
the PE/ground conductor connection of the socket and power plug match each other.
Connecting terminals: Connect the terminals according to the terminal connection
diagram. First connect the PE/ground conductor.
WARNING
Unsuitable cables, cable glands and / plug-in connectors
Risk of electric shock and material damage
Only use cable glands/plug connectors that meet the requirements for electrical safety.
Tighten the cable glands according to the instructions in the respective manuals / operating
instructions.
Close unused cable openings for electrical connections.
When replacing cable glands, only use ones of the same type.
Make sure the cables are tight after installation.
Note
Electrical circuit breaker in accordance with IEC 60947-1 and IEC 60947-3
In accordance with IEC 60947-1 "Standard for low-voltage switchgear and controlgear" and IEC
60947-3 "Switches, disconnectors, fuses" you require an electrical circuit breaker for the device.
We recommend commercially available automatic circuit breakers.
Connecting
5.1 Safety instructions
ULTRAMAT 23
68 Equipment Manual, 04/2022, A5E37100388-012