last two positions of each line are reserved for displaying certain analyzer statuses. The
meanings of these characters depend on the set language. The meanings are as follows:
Description Ger‐
man
Eng‐
lish
Frenc
h
Span‐
ish
Italian Polish
Maintenance request
(display lights up permanently)
A M D P R S
Fault present
(display lights up permanently)
S F F A E U
Limit violated
(display lights up permanently)
G L L L S O
Fault logged which is no longer present *
(display lights up permanently)
! ! ! ! ! !
Remote control
(display lights up permanently)
R R R R F Z
Function control:
• Analyzer is uncoded
• Access via RS485 serial interface and PRO‐
FIBUS interface
• AUTOCAL or warm-up phase running (dis‐
play ashes)
• Measured values for the inuence of inter‐
ference gas via ELAN not valid
F C C F C C
Pump running (display lights up permanently)
or
ow fault (display ashes)
P P P B P P
Analyzer is uncoded (display ashes) U U U D N K
* In the case of analyzers with an H
2
S probe, the following statuses can be displayed at the position for
the fault which is no longer present
Protection function of H
2
S probe running
(display lights up permanently)
H H H H H H
Protection function of H
2
S probe running, H
2
S
measured value is invalid (display ashes)
V V V V V V
Operation
7.3 Display and control panel
ULTRAMAT 23
90 Equipment Manual, 04/2022, A5E37100388-012