Operation
7.4 Operating modes
ULTRAMAT 23
102 Manual, 3/2016, A5E37100388-003
The <
CAL
> key cannot be used during the warm-up phase.
If the flow is too low during a zero calibration triggered by pressing the key, the analyzer
remains in this status until either the flow is sufficient or the zero calibration is aborted by
pressing the <
ESC
> key.
In addition to the <
CAL
> key, an AUTOCAL can also be triggered via the binary input. The
binary input has priority over the key.
7.4.3.5
The PUMP key
If the analyzer is equipped with an internal sample gas pump, this can be switched on and
off using the <
PUMP
> key. If the pump is switched off while the analyzer is in input mode, it
is also switched on again by pressing the <
MEAS
> key if parameterized accordingly (see
Configuration: Inputs/outputs/pump: Pump at CAL/MEAS (Page 142)).
In addition to use of the <
PUMP
> key, the pump can also be switched on and off via the
binary input. The binary input has priority over the key.