Functions
8.1 Diagnostics
ULTRAMAT 23
110 Manual, 3/2016, A5E37100388-003
–
The reference voltage for the electronics of
the analyzer.
–
: The supply voltage to the measuring bridge.
The temperature which determines the display contrast.
Setting of the LCD contrast is described in section
Parameter: Pump/LCD contrast (Page 130).
The actual value of the output current is displayed in mA for
each of the measured components (the dimension is not
shown for space reasons) as well as the start-of-
(either 0, 2 or 4 mA) and the full-scale value (20 mA) of the
output current range. Refer to section
Configuration:
Inputs/outputs/pump: Analog outputs (Page 134) for
adjustment of the start-of-scale values.
Analyzer status: Factory settings hardware
are parameters which are already set on de-
Serial No.
Revision
Order No.
The hardware configuration and release version can be read
here.
Analyzer status: Factory settings software
Factory settings are parameters which are already set on d
e-
Software/firmware release version
The software release version can be
read here.