Functions
8.3 Parameter
ULTRAMAT 23
126 Manual, 3/2016, A5E37100388-003
Parameters: Measuring ranges: Setting measuring ranges
-scale values of the measuring ranges can be set in the
third and fourth lines of this dialog. They must lie within the
factory settings, i.e. if an analyzer is facto
ry-set for a total
mg/m
3
, modifications are only possible
within this range. The following definitions also apply:
The smaller MR must not be greater than the higher MR.
The following input limits apply to the ranges:
– Lower limit: 0.01 times the smaller MR according to
factory setting (label)
– Upper limit: 1.1 times the higher MR according to factory
setting (label)
Smallest MR 1: 0 to 4 mg/m
3
Highest MR 2: 4 to 2200 mg/m
3