Functions
8.3 Parameter
ULTRAMAT 23
130 Manual, 3/2016, A5E37100388-003
Figure 8-1 Time constants
The following values are possible for the time constants:
T
90
within: 0.1 to 99.9 (s)
T
90
outside: 0.1 to 99.9 (s)
Threshold (data in %): 0 ... 100% of smallest range
8.3.5
Parameter: Pump/LCD contrast
You can use this menu to select two dialogs via which the
pump capacity and the contrast of the LCD can be changed.
8.3.5.1
Parameters: Pump/LCD contrast: Pump
In this menu you can:
Increase the pump capacity using the <
> or <
> key
Decrease the pump capacity using the <
> key
Store the set pump capacity using the <
ENTER
> key
Cancel the input using the <
ESC
> key.
Changes to the pump capacity are shown on the flowmeter
and directly in the menu display by the message "o.k.
" or "
not
o.k.".