Functions
8.3 Parameter
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
129
Parameters: Limits: H2S sensor protection
2
S' is selected as the component in the higher-level menu
2
S sensor protection' function is selected in addi-
tion to the limits. The 'Limit' function is described
in section
(Page 128).
In the 3rd line you can select the 'H
2
S sensor protection' func-
The adjacent menu display appears when you select the fun
c-
tion in which you can switch the 'H
2
S sensor protection' func-
2
S) is a corrosive gas, especially in wet
condition and in combination with other gases. This function
prevents damage occurring to the H
2
S sensor from high con-
2
S. This function is enabled by default (
).
Execution of this function is described in section
Probe protec-
(Page 156).
Parameters: Time constants
You can use this function to set various time constants to su
p-
press noise in the measured signal. During processing of the
measured signals, these time constants reduce the noise by
de
laying the signal. The time constant "
90
is effective
within an adjustable interval whose threshold values are d
e-
fined as a percentage of the smallest measuring range. The
time constant dampens small changes in signal (e.g. noise),
but becomes imm
ediately ineffective if a fast change in signal
exceeds a threshold. If the threshold is exceeded, the signal is
dampened by the time constant "T
90
" until it falls below
the threshold value again. "T
90
" then becomes effective