Description
3.4 Technical specifications
ULTRAMAT 23
42 Manual, 3/2016, A5E37100388-003
3.4.6
Hydrogen sulfide sensor
H
2
S sensor for measuring range from 5 to 50 vpm H
2
S
General information
Measuring ranges
Smallest measuring range
Largest measuring range
0 ... 5 vpm
0 ... 50 vpm
Service life of the sensor
Approx. 12 months
Operation mode Continuous measurement between 0 and 12.5 vpm
Discontinuous measurement between 12.5 and 50 vpm
AUTOCAL Cyclic adjustment (see H2S sensor with 'small' measuring range
(Page 161))
Climatic conditions
Permissible ambient temperature
During operation
During transportation and storage
+5 ... +40 °C (41 ... 104 °F)
-10 ... +55 °C (14 ... 131 °F)
Permissible ambient pressure
750 ... 1200 hPa absolute
Maximum operating altitude
2000 m above sea level
Influencing variables
Associated gases The hydrogen sulfide sensor cannot be used if the associated gas con-
tains the following components:
Compounds containing chlorine
Compounds containing fluorine
Heavy metals
Aerosols
Alkaline components (e.g. NH
3
>5 mg/m
3
)
Cross-inference (interfering gases) 1360 vpm SO
2
result in a cross-interference of <20 vpm H
2
S,
180 vpm NO result in a cross-interference of <150 vpm H
2
S,
no cross-interference of CH
4
, CO
2
and H
2
(1000 vpm)
Drift
< 1% per month
Temperature
< 3%/10 K referred to full-scale value
Air pressure
< 0.2 % of measured value per 1% pressure change
Note
Measuring ranges
The exact specification of the largest and smallest H
2
S ranges can be found on the label!