Functions
8.1 Diagnostics
ULTRAMAT 23
108 Manual, 3/2016, A5E37100388-003
Raw values
are the measured values in % of the full-scale
value (= 100 %).
Not limited concen.
are the measured values as they are
also displayed in measuring mode. You can approximately
display the concentrations here even with an upward or
downward violation of the largest range. Negative values
are also displayed (live zero). The current measuring range
is output on this display in the last two columns.
8.1.2.2
Analyzer status: Diagnostics values: (Electrochemical) O2 sensor
This dialog displays the diagnostics values of the electroche
m-
ical oxygen sensor (option). Meaning:
Sensorsig
is the current pressure-compensated voltage of
the O
2
sensor in mV
O
2
is the current oxygen value. Negative values are also
possible here
Inst. date
is the installation date of the O
2
sensor (see
section
Calibration: O2 measuring range: Sensor inst. date
(Page 114))
Inst.volt
is the pressure-compensated voltage of the O
2
sensor when it was installed.
8.1.2.3
Analyzer status: Diagnostics values: (Paramagnetic) O2 sensor
This dialog displays the diagnostics values of the paramagne
t-
ic oxygen sensor (option). Meaning:
Sensorsig
is the current voltage of the O
2
sensor in mV
O
2
is the current oxygen value. Negative values are also
possible here