Functions
8.2 Calibration
ULTRAMAT 23
118 Manual, 3/2016, A5E37100388-003
You can use this function to adjust the setpoint of the calibration gas for calibration of the
measuring range.
If the adjacent display appears, adjust the setpoint as follows:
Position the cursor at the beginning of the 2nd line (Setpoin
t)
key.
The adjacent display appears.
Now enter the desired setpoint using the arrow keys, and
then press the <
> key.
Exit the menu by pressing the <
> key.
You first need to select the H
2
S sensor in order to adjust it. To do this, navigate in the
operator menu as follows:
Calibration -> Please enter code -> Calibrate sensor -> Select sensor -> H
2
S sensor.
The adjacent display appears.
2
S sensor as described in the follow-
e following sequence must be observed when
Enter correction factors for temperature compensation of
zero point
Enter correction factors for temperature compensation of
sensitivity
Caibrate zero point of the H
2
S sensor
Caibrate measuring range of the H
2
S sensor
Enter installation date.
Steps 1, 2 and 5 are only carried out following installation of a new sensor.
The deflection signal of the sensor is subject to drift. This drift can only be detected through
regular checking and corrected as necessary. This involves using a calibration gas with a
defined concentration of hydrogen sulfide. We recommend monthly calibration with a
calibration gas to keep potential measurement uncertainty within strict limits: The hydrogen
sulfide concentration of this calibration gas should correspond to the concentration of the
sample gas, having a concentration of at least 10% of the largest full-scale value.