Area of application
Overview
Up to 4 gas components can be measured continuously and simultaneously with the
ULTRAMAT 23 gas analyzer. The analyzer has an infrared detector for infrared-sensitive
gases such as CO, N
2
O or CH
4
, and can be optionally fitted with up to two further
electrochemical sensors for O
2
and H
2
S as well as a paramagnetic O
2
sensor. The following
combinations are thus possible:
Table 3- 1 Possible combinations of ULTRAMAT 23
Electrochemical O
2
Paramagnetic O
2
sen-
H
2
S sensor
1 IR component +
x
--
x
2 IR components +
x
--
x
3 IR components +
x
--
--
Use of sensor is possible: x
Use of sensor is not possible: --
These combinations are available for a 19" rack unit with hosed gas paths. The following
deviations apply to other analyzer versions:
● Only IR components are available for 19" rack units with piped gas paths
● H
2
S sensors and O
2
sensors are available for bench-top units