Configuration: Special functions: Changing the codes/language
In the first two lines of this dialog, you can change the codes of
the two code levels 1 and 2 (see also section
Code levels
(Page
97)).
The factory settings for the two code levels are:
Code level 1:
111
Code level 2:
222
You can also reduce the number of code levels by assigning the
same
code to both levels.
With the introduction of firmware version 2.15.06, you can now
use letters and special characters for the code in addition to
numbers. With all older firmware versions, you could only enter
numbers.
The modified codes are effective imm
ediately. You should
therefore make a note of the changed codes and keep this at a
safe place.
In third line of this dialog you can change the language of the
input dialogs. The analyzer is designed for the following la
n-
guages:
German
English
Spanish
French
Italian
Polish
A change is immediately effective when you leave this dialog.