Functions
8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
145
Configuration: Special functions: ELAN/PROFIBUS/external interference: ELAN parameters
You can use this dialog to set the parameters for an ELAN ne
t-
You must set the channel addresses for the analyzers
present in an ELAN network. Addresses from 1 to 12 can be
set, where each address may only be used once.
You can set an analyzer name here. During communication
with ELAN, it can be used for the plain text identification of
the analyzers. An analyzer name may have up to 10
alphanumeric characters.
(ON/OFF)
The automatic transmission of measured values can be
switched on/off. With 'ON', the analyzer sends a measured
value frame cyclically every 500 ms.
To avoid considerably hindering communication within an
ELAN network, this function should only be set to 'ON' when
required (e.g. with correction of cross-interference).
(ON/OFF)
This function is not yet available. Therefore only
' is the
currently valid value.
Note
For further details on operation of the analyzer in an ELAN network, please refer to the
ELAN interface description (C79000
-B5274-C176 German/English).
Configuration: Special functions: ELAN/PROFIBUS/external interference: ELAN external interference
This function can be used for a correction calculation by mea
s-
uring the influence of an interfering gas b
y means of another
analyzer connected in the ELAN network.