Functions
8.4 Configuration
ULTRAMAT 23
148 Manual, 3/2016, A5E37100388-003
Configuration: Special functions: ELAN/PROFIBUS/external interference: PROFIBUS parameters
You can use this function t
o set the following PROFIBUS pa-
This function can be used to set a PROFIBUS station
address to all numerical values between 0 and 126.
This parameter is used to set the configuration response of
the device. The values 0, 1 and 3 can be set as valid
parameters. They have the following meanings:
–
:
Only the 'Profile ID number' is positively acknowledged
–
:
Only the device-specific 'ID number' is positively
acknowledged.
:
In order to work with the provided GSD and DD, the '
' parameter must have the value 1.
–
:
Only the 'Profile ID number' for multi-variable devices
(complex analyzers) is positively acknowledged.
You can use this function to enable the 8 relays of the add-
on card for control via PROFIBUS. To allow activation, none
of these relays must already be occupied by a device-
internal function.
:
The function 'PB relay' is only possible as of PROFIBUS
card firmware version 2.0.0 (shown as Firmware in the
figure).
If the 'Diagnosis' parameter is selected, the 'Firmware'
display appears with, for example, the following parameters:
–
The firmware version is displayed here
–
The version of the boot firmware is displayed here
–
The name assigned to this analyzer in the network (or
the first 16 characters).