Functions
8.4 Configuration
ULTRAMAT 23
150 Manual, 3/2016, A5E37100388-003
Configuration: Special functions: Factory data/reset/units: Reset
You can use this function to restart the analyzer.. When you
select this function, you will be warned (see
adjacent display) that the analyzer initially runs
through a warm-
up phase following the restart and is thus unavailable for mea
s-
urements for a certain time.
> key initiates the restart with the warm-up phase.
Triggering of the restart can
be prevented here using the
Configuration: Special functions: Factory data/reset/units: Change units
In the second line of this menu display, you ca
n change the
-set units of the measured components.
After changing the unit, the display of the 'MB min
' and '
' parameters is adapted accordingly.
This dialog is specific to the component.
-scale values may assume unusual values as a result of the component-specific
conversion factors. Subsequent adaptations can be carried out as
described in
Parameters: Measuring ranges: Setting measuring ranges (Page 126). In addition,
you should also check these parameters follow
ing this change:
Setpoints of the calibration gases (section Calibration: Infrared measuring range: Set
span gas values (Page 113))
Settings for the limits (section Parameters: Limits (Page 128)).