EasyManua.ls Logo

Siemens ULTRAMAT 23 - Page 21

Siemens ULTRAMAT 23
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Description
3.1 Area of application
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
21
Further applications:
Environmental protection
Chemical plants
Cement industry
Special versions
The ULTRAMAT 23 with 2 IR components without pump is also available with two
separate gas paths. This allows the measurement of two measuring points as used e.g.
for the NO
x
measurement before and after the NO
x
converter.
The ULTRAMAT 23 gas analyzer can be used in emission measuring systems and for
process and safety monitoring.
For measurement of CO, NO, SO
2
and O
2
according to 13th BlmSchV and TA Luft, TÜV-
approved versions of the ULTRAMAT 23 are available.
The analyzers of the 7MB2355, 7MB2357 and 7MB2358 series are suitability-tested in
accordance with EN 15267 (emission measurements).
The current certificates and approvals can be found on the Internet at: Approvals and
certificates (https://support.industry.siemens.com/cs/de/en/ps/17728
)
Version with faster response time:
There is no connection between the two condensation traps, so that the complete sample
gas flow passes through the detector (only 1/3 of the flow in the normal version of the
analyzers), i.e. the response time is 2/3 faster. The functions of all other components
remain unchanged.
Version with chopper section purging:
This version consumes approx. 100 ml/min of purging gas; you must set an inlet pressure
of 300 kPa (3 bar).
Variant with specially cleaned gas path: This variant is envisaged for the applications
requiring 'Cleaned for O
2
service'.
Benefits
AUTOCAL can be carried out with ambient air (dependent on the measured component)
and is therefore highly cost effective because calibration gases and accessories are not
required
High selectivity thanks to multi-layer detectors,
low cross-sensitivity to water vapor
Sample chambers can be cleaned (dependent on the version), resulting in cost savings
through reuse following contamination
Menu-assisted operation in plain text, thus high operational safety
Service information and logbook, cost savings through preventive maintenance and help
for service and maintenance personnel
Increased safety through coded operator levels, thus protection against unauthorized
access or clumsy working