Page 27 of 92
Chapter: 2. Operation
For calibration, see: “5. Calibration procedure” on page 51.
The calibration screen is default set to "Disabled". To enable calibration please see "Enabling of calibration"
on page 52.
Note: If the system is not calibrated, a warning "NOT CALIBRATED WT BT" will ash on the lower left
part of the screen.
Screen calibration, Menu 1, Auto calibration.
Softkey Name Range/value Default
value
Description Activate
with hidden
button
1 MENU 1 - 4 1 Menu 1 is selected.
2 START Leg 1 start/nish
Leg 2 start/nish
Leg 1 Used to start/nish test trip. Yes (1 beep)
3 CALIBR Activate Used to move data, collected during test trip into
calibration table. Note: Active when calibration
is within limits
4 TRIPS LIST 1 - 5 1 Used to list between dierent test trip data
sets. Only one test trip is displayed on the
screen at a time.
5 TRIP BT On/O O
6 CALIBR DIS 0.054 - 5.400 NM 1.000 NM Used to adjust length of the test leg.