6
Doses
e pump can deliver two types of doses, an Automatic Dose
(programmed to deliver a specic dose at a preprogrammed
time) as well as Demand Dose (delivered by the patient as
needed). You can also program lockout times so that you can
control the time between doses.
Other features
e pump has a large display (or screen) where all program-
ming, operating and alarm information is displayed. Pro-
gramming of the pump is menu-driven, like an ATM or cell
phone.
e pump is powered by one AAA (IEC LR03) alkaline bat-
tery, which is readily available at most grocery stores, hard-
ware stores, drug stores, and electronic stores. e expected
battery life is approximately 2 weeks (battery low alert) at
0.124 ml/hr.
Caution: Do not use NiCd, nickel metal hydride, carbon
zinc (heavy duty), lithium or any rechargeable batteries.
ey will not power the pump properly, and the battery
life indicator on the home screen may not show the correct
amount.
As long as the pumps labels and outer shell are intact, the
pump is watertight to a depth of 8 feet (2.4 meters) for
30 minutes or 12 feet (3.6 meters) for 3 minutes.
Introduction
e CADD-MS
3 Ambulatory Infusion Pump provides
measured medication therapy to patients in hospital or out-
patient settings. Any medication therapy must be overseen by
a physician or certied, licensed healthcare professional.
Indications
e CADD-MS
3 pump is a syringe-based ambulatory infu-
sion pump designed for subcutaneous, intravenous, epidural
and intrathecal infusion of medication.
Contraindications
e pump is not indicated for anyone who cannot follow the
instructions for use or perform basic troubleshooting and
maintenance activities associated with ambulatory pump use.
Delivery features
e pump can be used to deliver medication in two ways, as a
continuous rate and as an added dose.
Continuous Rate
e pump can be programmed to deliver a steady ow of
medication called the Continuous Rate. You can program
up to 48 time/rate segments per 24 hour period to meet the
patient’s medication needs throughout the day and night.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals